Users Manual Part 1
Clinical User’s Guide
IBA | 4-1 |
Volume 1 - Treatment Session
Chapter 4
• • • • • •
Miscellaneous Safety Hazards
Cleaning, a Basic Safety Requirement
CAUTION The Proton Therapy System (PTS) parts which are not in direct contact
with the patient can be cleaned with the following products:
* Isopropyl Alcohol - Concentration up to 70 %
* Ethyl Alcohol - Concentration up to 90 %
* Bleach solution - up to 100 ppm active Chlorine
* Formaldehyde - up to 35 %
* Didecyldimethylammonium chloride - up to 3 mg/g
The following exclusions apply:
* Couch: you must refer strictly to the user instructions of the couch.
* Robot Patient Positioning System (PPS): you should avoid bleach and
other chlorine solutions for Robot PPS cleaning.
During the cleaning process please ensure to:
* Avoid water to penetrate inside equipment parts.
* Avoid using sprays such as the cleaning product penetrates parts.