Use Instructions
52 | International regulations
21 International regulations
The Invia Ease pump is intended to be used in the professional healthcare facility environment
and home healthcare environment and is EMC tested in conformity with the requirements of
the standard IEC 60601-1-2:2014 4th Edition according to clause 7 and 8.9.
The pump needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in these instructions for use. Portable and
mobile RF communications can affect the pump.
21.1 Electromagnetic compatibility (EMC)
NOTICE
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reason-
able protection against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the distance between the equipment and receiver.
• Connect the equipment to an outlet on a circuit different from that to which the receiver
is connected.
• Consult the dealer or an experienced radio/TV technician for help.
To prevent malfunction resulting from electromagnetic interference:
• The Invia Ease pump should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Invia Ease pump should be observed to verify
normal operation in the configuration in which it will be used.
• Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies, RFID can affect
the Invia Ease pump and should be kept at a distance of at least 30 cm away from the
device.
• Use of accessories or cables other than those provided by the manufacturer of this device
could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
WARNING