Use Instructions
EN
Alarms and troubleshooting | 41
Alarm table
To close the instructions after acknowledging the alarm and to return to the Main screen
click
Alarm type Potential cause Troubleshooting
instructions
Notes
Battery low
• Battery charge has
fallen to 20 %
• Alarm will repeat if
the battery charge
drops to 15 % and
subsequently to 10 %
Charge the pump
battery
For instructions,
refer to chapter
“Battery status
and charging” on
page 20.
If battery charge
drops to 5 %, it will
lead to the “Battery
empty” alarm and
therapy will be auto-
matically paused.
Temperature high
The pump is next to a
heat source or is under
a cover
Remove the pump
from the heat source
or uncover the pump
The pump is at risk
of overheating. If the
pump is not cooled, it
will lead to “Tempera-
ture exceeded” alarm
and therapy will be
automatically paused.
No canister detected
The pump does not
detect a canister
• Attach a canister to
the pump
• Release and
reposition
the canister if
needed
If a canister is missing
or is not correctly
attached, it will lead to
“High leakage” alarm.
Pump in wrong
position
The pump is positio-
ned on its side or is
upside down
Place the pump
in the upright position
If the pump is in a
wrong position for too
long, it could lead to
“Canister full” alarm.
i
“Low priority” alarms:
– Therapy continues to run
– Status indicator illuminates yellow
NOTICE
”Low Priority” alarms that are not resolved will lead to “Medium Priority” alarms.