Use Instructions

EN
Alarms and troubleshooting | 41
Alarm table
To close the instructions after acknowledging the alarm and to return to the Main screen
click
Alarm type Potential cause Troubleshooting
instructions
Notes
Battery low
Battery charge has
fallen to 20 %
Alarm will repeat if
the battery charge
drops to 15 % and
subsequently to 10 %
Charge the pump
battery
For instructions,
refer to chapter
“Battery status
and charging” on
page 20.
If battery charge
drops to 5 %, it will
lead to the “Battery
empty” alarm and
therapy will be auto-
matically paused.
Temperature high
The pump is next to a
heat source or is under
a cover
Remove the pump
from the heat source
or uncover the pump
The pump is at risk
of overheating. If the
pump is not cooled, it
will lead to “Tempera-
ture exceeded” alarm
and therapy will be
automatically paused.
No canister detected
The pump does not
detect a canister
Attach a canister to
the pump
Release and
reposition
the canister if
needed
If a canister is missing
or is not correctly
attached, it will lead to
“High leakage” alarm.
Pump in wrong
position
The pump is positio-
ned on its side or is
upside down
Place the pump
in the upright position
If the pump is in a
wrong position for too
long, it could lead to
“Canister full” alarm.
i
“Low priority” alarms:
– Therapy continues to run
– Status indicator illuminates yellow
NOTICE
”Low Priority” alarms that are not resolved will lead to “Medium Priority” alarms.