Use Instructions
EN
Change pump settings | 37
11.3 Alarm volume
Select the sound of the acoustic alarm. The sound always starts low and continues to get
louder until it reaches its maximum volume. The default alarm volume setting is Medium.
11.5 Language
1. Select the language from the list.
2. Click to return to the pump settings menu.
11.6 Bluetooth
– Bluetooth functionality is presently inactive regardless of Bluetooth setting (On or Off).
– Default setting of the Invia Ease pump is Bluetooth Off.
– Bluetooth interface is used by the Medela service centers only.
11.3 Vacuum intensity
Select the speed at which the negative pressure is applied and the dressing compresses. The
Low vacuum intensity requires the longest time to apply the target negative pressure while
the High vacuum intensity is the quickest to apply the target negative pressure at the
dressing. The default vacuum intensity setting is Medium.
The Invia Ease pump can be reset to the default factory settings:
– Therapy setting is constant mode in –125 mmHg
– Language is English
– Clinician access is Disclaimer
– Alarm volume is medium
– Vacuum intensity is medium
– Unit pressure is mmHg
– Bluetooth is Off
11.7 Factory settings reset