Use Instructions

26 | Operating instructions
Therapy interruption: The recommended application of the therapy with Invia Ease
NPWT system is 24 hours without interruption. If therapy is interrupted for more than
2 consecutive hours (120 minutes), the dressing should be changed and therapy restarted
by a clinician in order to avoid the risk of infection or sepsis.
WARNING
8.9 Pause therapy
The pump may be paused for a variety of reasons including:
To change the wound dressing. For detailed information regarding dressing applications,
consult the appropriate Invia dressing Instructions for Use.
To change the cannister.
To temporarily detach the patient from the pump.
The pump allows the therapy to be paused for up to 30 minutes. After therapy has been
paused for 30 minutes, the “Pause reminder” alarm will appear on the touch screen.
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Pause
reminder” on page
43.
i
Click and hold for 3 seconds.
Pause screen will appear, the pump motor will stop working and the light indicator will switch
from green to yellow.
Pause screen
Open instructions
Pause therapy timer
Log dressing change
Restart therapy
Access Clinician menu
To pause therapy: