Use Instructions

20 | Operating instructions
8 Operating instructions
Never operate the pump if it is not working properly, if it has been damaged or dropped
into water. If damage is found, contact the Medela Customer Service or the system
provider.
WARNING
The Invia Ease NPWT system is intended to be used by clinicians or by adequately trained
lay users (patients or caregivers).
8.1 Pump positioning
This product contains lithium-ion batteries, which bear risk of fire, explosion and burns.
Only use the charger that comes with the pump. An inadequate power source may result
in a fire hazard, electric shock or malfunction of the pump.
Prior to charging, always inspect the charger and the cord for damage. If damage is
found, immediately disconnect the charger and call the Medela Customer Service or
the system provider.
WARNING
8.2 Battery status and charging
NOTICE
Medela recommends continuous charging while the pump is running overnight to ensure a
full charge.
If the pump is not being used, the battery must be charged approximately once every
6 months to ensure optimum function.
The Invia Ease pump should remain in an upright position or can be placed on its front or
its back during use.
If the Invia Ease pump is placed in an inappropriate position (upside-down or on its side),
it will lead to the “Pump in the wrong position” alarm.
If the canister filter becomes clogged because the pump is in an inappropriate position
for too long, it will lead to the “Canister full” alarm. The canister must then be replaced.
i
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Pump in
the wrong position” on page
42.
i
For troubleshooting instructions refer to chapter “Alarms and troubleshooting/Canister
full” on page
44.