User's Manual

20
2.3 PREPARATION FOR AN REM TEST
The REM test must be conducted only by a person specialized in
and authorized to perform audiology procedures. Use of the
instrument by a person without the necessary expertise could result
in damage to the hearing of the patient.
Change the probe for each new patient. Reusing the tube on
different people could result in the spread of ear infections.
Do not push the tube too far into the patient's ear: there is the risk
of perforating the eardrum. Refer to the ruler marked on the
headset to determine the depth of insertion.
Always carry out an ear inspection before proceeding with the REM
test, to make certain there are no obstructions in the ear that could
affect the results of the examination.
2.3.1 Measurements in the ear
When preparing the patient for an REM test, observe the following
procedure:
1. Position the patient at around 1 metre from and facing the speaker
of the instrument, and in such a way that the ears are placed
approximately at the same height as the instrument.
2. Position the black ring of the tube so that the placement
corresponds to the required insertion distance (refer to the