Product Manual
Invacare®Platinum™Mobile
Surge
IEC61000-4-5
±1kVline(s)toline(s)
±2kVline(s)toground
±1kVline(s)toline(s)
±2kVline(s)toground
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.
Voltagedips
IEC61000-4-11
0%U
T
;1cycle
and
70%U
T
;25/30cycles
Singlephase:at0°
0%U
T
;1cycle
and
70%U
T
;25/30cycles
Singlephase:at0°
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.IftheuseroftheDevicerequires
continuedoperationduringpowermainsinterruptions,it
isrecommendedthattheDevicebepoweredfroman
un-interruptiblepowersupplyorabattery.
U
T
isthea.c.mainsvoltagepriortoapplicationofthe
testlevel.
VoltageInterruptions
IEC61000-4-11
0%U
T
;250/300cycle0%U
T
;250/300cycle
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.IftheuseroftheDevicerequires
continuedoperationduringpowermainsinterruptions,it
isrecommendedthattheDevicebepoweredfroman
un-interruptiblepowersupplyorabattery.
U
T
isthea.c.mainsvoltagepriortoapplicationofthe
testlevel.
Powerfrequency
(50/60Hz)magnetic
field
IEC61000-4-8
30A/m30A/m
Powerfrequencymagneticfieldsshouldbeatlevels
characteristicofatypicallocationinatypicalcommercial
orhospitalenvironment.
661194969-A-00