Product Manual
TechnicalData
8.4ElectromagneticCompliance(EMC)
Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheDeviceshouldassure
thatitisusedinsuchanenvironment.
Emissionstest
Compliance
Electromagneticenvironment-guidance
RFemissions
CISPR11
GroupI
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsarevery
lowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.
RFemissions
CISPR11
ClassB
Harmonicemissions
IEC61000-3-2
ClassA
Voltagefluctuations/
flickeremissions
IEC61000-3-3
Complies
TheDeviceissuitableforuseinallestablishmentsincludingdomesticestablishmentsandthose
directlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildings
usedfordomesticpurposes.
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheDeviceshouldassure
thatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevel
Electromagneticenvironment–guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±8kVcontact
±15kVair
±8kVcontact
±15kVair
Floorsshouldbewood,concreteorceramictile.Iffloors
arecoveredwithsyntheticmaterial,therelativehumidity
shouldbeatleast30%.
ElectricalFast
transient/burst
IEC61000-4-4
±2kVforpowersupply
lines
±1kVforinput/outputlines
±2kVforpowersupply
lines
±1kVforinput/outputlines
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.
1194969-A-0065