Product Manual
Invacare®Platinum™Mobile
DeliveredOxygenPulseVolumes:
•ThenominalpulsevolumespublishedinthefollowingtableisinmillilitersatSTPD(standardtemperatureandpressuredry)conditionsand
applyovertheratedrangesforambienttemperature,humidity,andatmosphericpressure.
•Maximumvariationfromnominal:+/-15%
4ftto25ftCannula/OxygenTubingLengths
BreathsPerMinute
152025303540
PulseSetting=P1
14.711.08.87.36.3
5.5
PulseSetting=P2
29.322.017.614.712.611.0
PulseSetting=P3
44.033.026.422.018.916.5
PulseSetting=P4
58.744.035.229.325.122.0
8.3RegulatoryListing
ETLcertified
complyingwith:
EN/IEC60601-1;Ed:3.1
EN/IEC60601-1-2;Ed:4
AAMIES60601–1(UnitedStates)
CSA22.2No.60601–1(Canada)
ISO80601–2–69
ISO80601–2–67
IEC60601–1–6
IEC60601–1–8
IEC60601–1–11
RTCADO160G
641194969-A-00