Product Manual

Invacare®Platinum™Mobile
ConserverTrigger
Sensitivity:
< 0.18cmH
2
Opressuredrop(Forallcannulalengths)
Factoryset—noadjustment,pressureactivated
Onlypatientrespiratoryeffortsthatachievethetriggerpressurewillresultinthedeliveryofanoxygenbolus.
ConserverBreathRate
Capacity:
15–40BPM(breathsperminute)withoutreductionofbolusminutevolume
MaximumOutletPressure:28.5psig(197kPa)
CannulaRequirements:Length:4–25ft(1.2–7.6m)includingalloxygentubing
Tubing:crush-proof,singlelumen
Adult,standardflow(ratedforupto6L/mincontinuousflow)forlengthsupto7ft)
Adult,highflow(ratedforupto15L/mincontinuousflow)forlengthsgreaterthan7ftto25ft)
ExampleofPossibleCannulaModel:WestmedInc.PartNumber0194(4ftlength)
BatterySpecifications
(eachbattery):
Rechargeablelithium-ion,14.4V,5800mAh,83.5Wh,500fullcharge/dischargecyclelife
BatteryShelfLife:12monthsfromdateofmanufacture
ConditionOneBattery
TwoBatteries
PulseSettingP1
5hr5min
10hr10min
PulseSettingP23hr30min7hr0min
PulseSettingP32hr20min4hr40min
BatteryDuration:
(Timesareapproximate)
PulseSettingP41hr45min3hr30min
621194969-A-00