Product Manual
Invacare®Platinum™Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryCold
INDICATORS:
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
Batteryistoocoldtoallowthebatteryto
continueoperating.
Concentratorfunctionhasshutdown.
1.Moveconcentratortowarmer
surroundings.Allowconcentratorto
warmupto41°F(5°C),OR
2.UseACorDCpowerandremove
battery,OR
3.Insertanalternatechargedbatteryand
removethecoldbattery,OR
4.Changetoanothersourceofoxygenifno
otherpowersourceisavailable.
NOTES:
Toclearthealarm,turntheconcentratoroff.Completesolutions1,2,3,or4beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
OxygenFault
INDICATORS:
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
Theoxygenoutputpurityhasfallentoavalue
below73%.
Concentratorfunctionhasshutdown.
1.Verifyconcentratorisinrecommended
environmentaltemperature.Referto
SpecificationsinTechnicalData.
2.Cleanintakefilterandensurebothintake
andexhaustarenotblocked.
3.Changetoanothersourceofoxygenif
alarmcontinues.
NOTES:
Toclearthealarm,turntheconcentratoroff.CompleteSolution1and/or2beforerestarting.Thealarmlimithasatoleranceof+/-2%
Alarmsignalgenerationmaybedelayedbyupto128secondsduetothemonitoringalgorithmusedtopreventnuisancealarms.
Callyourproviderandreporttheproblemifalarmcontinues.