Product Manual

Invacare®Platinum™Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryDepleted
INDICATORS:
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
Batterygauge:
Nosegmentsilluminated
Remainingbatterycapacityhasbeen
completelyused.Batteryrequirescharging.
Concentratorfunctionhasshutdown.
1.Connectoxygenconcentratortoeither
theACorDCpower,OR
2.Insertanalternatechargedbattery.
Removethedepletedbatteryandcharge
withtheoptionalexternalbatterycharger
accessoryforfutureuse,OR
3.Ifnootherpowersourceisavailable,
changetoanothersourceofoxygen.
NOTES:
Toclearthealarm,theconcentratorneedstobeturnedoff.Completesolution1or2beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
UnitTooHot
INDICATORS:
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
Concentratoristoohottoallowittocontinue
operating.
Concentratorfunctionhasshutdown.
Internalfanwillcontinuetoruntohelplower
internaltemperature.
1.Moveconcentratortocooler
surroundings.Allowconcentrator
tocooldownto104°F(40°C).
2.UseACorDCadapterforpower.
3.Cleanintakefilterscreen.
4.Changetoanothersourceofoxygen
whilewaiting.
NOTES:
Toclearthealarm,turntheconcentratoroff.Completesolutions1,2,and/or3beforerestarting.