Product Manual

Invacare®Platinum™Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowOxygen
SeeManual
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Theoxygenoutputpurityhasfallentoavalue
between73%and87%.
1.Verifyconcentratorisinrecommended
environmentaltemperature.Referto
SpecificationsinTechnicalData.
2.Cleanintakefilterscreenandensureboth
intakeandexhaustarenotblocked.
3.Turnconcentratoroff,thenonagainto
retry.
NOTES:
Thealarmwillturnonwhentheoxygenpurityhasdroppedbelow85%.
Thealarmwillturnoffwhentheoxygenoutputpurityhasincreasedaboveto87%.
Thealarmlimitshaveatoleranceof+/-2%.
Alarmsignalgenerationmaybedelayedbyupto128secondsduetothemonitoringalgorithmusedtopreventnuisancealarms.
Callyourproviderandreporttheproblemifthewarningcontinueswithnextuse.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Cannotcharge
Allowtocool
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryistoohottoallowthebatteryto
charge.
ThisalarmonlyoccurswhenExternalPower
isconnected.
1.Moveconcentratortocooler
surroundings.Allowconcentrator
tocooldowntolessthan104°F(40°C),
OR
2.Removebattery(withACorDCpower
connected)toallowbatterytocool.
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreached,whenthebatteryisremoved,orwhenexternalpowerisdisconnected.
521194969-A-00