Product Manual

Invacare®Platinum™Mobile
6.3WearandTear
Invacarereservestherighttoaskforanyitembackthathasan
allegeddefectinworkmanship.SeeWarrantythatshippedwiththe
productforspecificwarrantyinformation.
RefertoMaintenanceinthismanualforproperpreventative
maintenancescheduleanduseoftheproduct.
Thisisonlyageneralguidelineanddoesnotincludeitemsdamaged
duetoabuseandmisuse.
Normalwearandtearitemsandcomponentsforthisproductare
listedbelow.
Description
ExpectedLife
Compressorinletfilter1,460hours
Outletfilter730hours
Carrybag18months
Battery
18months
Sievebedassembly1year
Compressor4,380hours
Allvalves4,380hours
Expectedlifevaluesshownashoursisinreferencetohours
ofconcentratoroperation.RefertoViewingHourMeterin
Maintenance.Expectedlifevaluesshownasmonthsoryears
isinreferencetodurationoftimesincethedateofpurchase.
Sieveisaporousfilteringmaterialandisconsideredawear
item.Somefactorsthatcouldaffectsievemateriallifeinclude
humidity,temperature,particulates,aircontaminates,air
intake,vibrationandotherenvironmentalconditions.
44
1194969-A-00