4.6WarmUpPeriod
CAUTION!
RiskofMinorInjuryorDiscomfort
Duringthewarmupperiod(typicallylessthanfive
minutes)oxygenoutputisnotwithinspecificationslisted
inSpecificationsinTechnicalData.
Concentratormaybeusedduringwarmupperiod.
Afterturningontheconcentratorawarmupperiodofupto15
minutesisneededtoreachthespecifiedoxygenpurity.“Warming
Up”isshownonthedisplayscreenduringthisperiod.
Whilewarmingup,theconcentratorwillautomaticallydeliverapulse
ofoxygenabouteveryfoursecondswhennobreathisdetected.
Thenasalcannulashouldbeconnectedtotheconcentratorduring
warmupperiod.
Whentheconcentratorhasfinishedwarmingup,thecurrentpulse
settingwillbeshownonthedisplayscreen.Ifnobreathisdetected
afterwarmuphasended,theNoBreathDetectedalarmwilloccur.
RefertoAlarmConditionsinTroubleshooting.
4.7BreathingwiththeUseoftheConcentrator
Asyoubreathewiththenasalcannulainstalled,apulseofoxygenis
deliveredeachtimetheconcentratorsensesaninhalation.Refer
toConnecting/PositioningtheNasalCannulainSetupforcannula
installationinstructions.
Ifnobreathisdetectedfor15seconds,theNoBreathDetected
alarmwilloccur.RefertoAlarmConditionsinTroubleshooting.
4.8ReadingtheDisplayScreen
PulseFlowSettingandPulseIcon
Aftertheconcentratorfinisheswarmingup,thecurrentoxygenpulse
flowsettingBandpulseiconCareshownonthedisplayscreenA.
Thepulseiconflashesoneachtimeanoxygenpulseistriggered.
Thepulseflowsettingandpulseiconarenotshownwhen
alarmtextisdisplayedonscreen.
InformationalText
Informationaltextisshownonthedisplayscreenduringcertain
periodsofconcentratoroperation:
“WarmingUp”displayedwhiletheconcentratoriswarming
upafterfirstbeingturnedon.
“PoweringDown”displayedwhiletheconcentratoris
poweringdownafterbeingturnedoff.