Oxygen Machine Operator's Manual IRC5LXO2, IRC5LX, IRC10LXO2, IRC10LX
SECTION 6—MAINTENANCE
Platinum
™
Series 34 Part No 1118353
WARNING
DO NOT overfill humidifier bottle.
FIGURE 6.2 Preventive Maintenance Record
ON EACH INSPECTION
MODEL NO. I R C SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
INVACARE PLATINUM OXYGEN CONCENTRATOR
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY 180 DAYS FOR IRC5LX (EVERY 90 DAYS FOR IRC10LX)
Check Oxygen Concentration
(UNITS WITHOUT SensO
2
)
ANNUALLY, DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Oxygen Concentration
(SensO
2
MODELS)
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*Refer to Preventive Maintenance section of Service Manual.