Product Manual

LABEL LOCATION
Part No 1100873 9 HomeFill
®
Compressor
LABEL LOCATION
DEALER INFORMATION
NOTE:Invacarerecommendsleavingafullcylinderofoxygenwith
thepatientaftersettinguptheHomeFillcompressor.
BeforeleavingaHomeFillCompressorwithauser,the
followingchecklistMUSTbecompleted:
Makesurethepressuretestoncylinderhasnotexpired.
Instructtheuseronthesafeoperationoftheconcentrator
andreviewallwarnings.
InstructtheuseronthesafeoperationoftheHomeFill
compressorandreviewallwarnings.
Leaveafullcylinderofoxygenwiththeuser.
Leaveacopyofallwarningsandfillinginstructionswith
theuser.
WARNING
Keep all connections free of oil and grease to avoid violent
ignition. See Owner’s Manual before cleaning.
P/N 1075220 Rev C - 10/04
WARNING
DO NOT REMOVE UNIT COVER
This unit contains moving and high pressure
components. Manufacturer Service Only.
P/N 1075221 Rev B- 10/04
WARNING
Keep all connections free of oil and grease.
See Owner’s Manual before cleaning. Never oil or lubricate
coupling.
Cylinders filled on this unit MUST be labeled “Contents of
this cylinder are the product of an oxygen concentrator.”
P/N 1075238 Rev C - 11/17/06