SECTION 5—COMPRESSOR OPERATION
HomeFill
®
Compressor 28 Part No 1100873
Turning the Compressor On
NOTE:Forthispr ocedur e,refertoFIGURE5.5onpage 29.
1. Makesuretheconcentratoris On.Refertotheconcentrator
OwnersManual.
CAUTION
TheconcentratorflowratetothepatientMU STbesetat
maximum:3L/minorlessforPlatinum5,XL,andPerfecto
2
and
5L/minforPlatinum 10.Otherwise,theoxygenleveltothe
compressorwillbe<90%andthecompressorwillnotfillthe
cylinder.
2. Settheconcentratorflowratetothepatientto3L/minorless
forPlatinum5,XL,andPerfecto2and5L/minforPlatinum
10.RefertotheconcentratorOwnersManual.
3. Makesurethecylinderisconnectedtothecompressor.Refer
toConnecting theCylindertotheCompressoronpage 23.
4. Pushthepow erswitchonthecontrolpaneltotheOn()
position.
5. Examinethecontrolpanel.Thefollowingsequenceofevents
shouldoccur:
A. 03minutes‐TheO
2
BelowNormal(YELLOW)lightwill
belitwhilethecompressorw armsup.
B. Afterthreeminutes‐TheFILLING(GREEN)lightwillbe
onwhilethecylinderisfilling.
C. TheFULL(GREEN)lightwillbeonwhenthecylinderis
finishedfilling.ProceedtoSTEP6.
6. Removethefullcylinder.RefertoDisconnectingthe
Cylinderfrom theCompressoron page 26.
7. Pushthepow erswitchtotheOff(O)position.