Product Manual

SECTION 5—COMPRESSOR OPERATION
Part No 1100873 23 HomeFill
®
Compressor
SECTION 5—COMPRESSOR
OPERATION
Connecting/Disconnecting the Cylinder
to/from the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthecylinder
andthe compressor.Otherwise,sev ereinjuryand/ordamage
mayoc cur.
Connecting the Cylinder to the Compressor
WARNING
DONOTdropoxygencylinders.Usetw ohandswhen
handling/transportingoxygencylinders.Otherwis e,injuryor
damagemayoccur.
DONOTusealiquidleakdetectortotestforleaks.
NOTE:Forthispr ocedur e,refertoFIGURE5.1onpage 24and
FIGURE5.2onpage 25.
1. Examinethecylindergauge.Ifthecylinder pressureisles s
than1500psig,proceedtoSTEP2.
NOTE:Forlocationofthecylindergauge,refertothecylinder’s
manual.
NOTE:If thecylinderpressur eisgreaterthan1500psig,DONOT
attempttotopoffthiscylinder.Itmaynotfill.
2. SettheregulatorflowdialonthecylindertoOff.Refertothe
cylindersmanual.
3. Removethecylinderandconnectorfillportcov ers(if
present).