User Manual Invacare® Element™ Pneumatic Conserver IOC100P Dealer: This manual MUST be given to the end user. User: Before using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment.
TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES ............................................... 4 Receiving Inspection ........................................................................... 4 Intended Use ........................................................................................ 4 Disposal Instructions .......................................................................... 4 Principles of Operation......................................................................
SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
SPECIAL NOTES Principles of Operation The oxygen conserving regulator is designed to be used with high pressure oxygen systems. It consists of a cylinder connection, cylinder contents gauge (if equipped), high to low pressure regulator, orifice plate and a conserving demand module. The regulator reduces the high pressure of the cylinder to the working pressure of the orifice plate. The orifice plate uses calibrated orifices to deliver a selected flow to the conserving demand module.
INTRODUCTION INTRODUCTION The oxygen conserving regulator includes a combination of low‐pressure regulator and an oxygen conserver, designed for use with a cylinder as an ambulatory oxygen system. It is capable of delivering a precise amount of supplemental oxygen at the optimal point in the breathing cycle. The conserver greatly increases the efficiency in the delivery of oxygen, maximizing the beneficial effects and eliminating unnecessary oxygen waste.
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES When using the oxygen conserving regulator, basic precautions should ALWAYS be followed, including the following: WARNING SECTION 1 ‐ GENERAL GUIDELINES contains important information for the safe operation and use of this product.
SECTION 1—GENERAL GUIDELINES Store product in a clean area when not in use. NEVER smoke in an area where oxygen is being administered. NEVER use aerosol sprays near the equipment. ALWAYS confirm prescribed dose before administering to patient and monitor on a frequent basis. DO NOT use if dirt or contaminants are present on or around cylinder, valve, product or connecting devices. Not suitable for use in the presence of flammable anesthetic mixture.
SECTION 1—GENERAL GUIDELINES DO NOT allow oil from your hands or other sources to come into contact with the cylinder post valve. These substances may become flammable in the presence of oxygen and cause injury. Avoid touching the cylinder post valve whenever possible. Example of items to avoid are hand creams or lotions, cooking oils, suntan oils or sun blocks and similar oily products. If you use these types of products, wash your hands prior to use.
SECTION 2—FEATURES SECTION 2—FEATURES DANGER This product is not intended as a life‐sustaining or life‐supporting device. CAUTION Missing or illegible labels MUST be replaced. Contact customer service at the number on the back page. Model IOC100P Cylinder Post Valve T-Handle Clamp Cylinder Alignment Pins CGA870 Seal Washer Oxygen Pressure Gauge w/ Rubber Guard Selector Switch Oxygen Supply Outlet Fitting Vent Hole FIGURE 2.
SECTION 2—FEATURES Selector Switch: Enables the user to select the desired setting, as well as “OFF” and “CF” (continuous flow). When not in use, the switch should be turned to the “OFF” position. Oxygen Supply Outlet: Use this fitting to attach a standard cannula. Cylinder Alignment Pins: When assembling the unit, these parallel pins MUST be inserted into the holes on the CGA870 post valve.
SECTION 3—INSTALLING THE CONSERVING REGULATOR SECTION 3—INSTALLING THE CONSERVING REGULATOR WARNING DO NOT allow oil from your hands or other sources to come into contact with the regulator gasket or cylinder post valve. These substances may become flammable in the presence of oxygen and cause injury. Avoid touching the cylinder post valve whenever possible. Example of items to avoid are hand creams or lotions, cooking oils, suntan oils or sun blocks and similar oily products.
SECTION 3—INSTALLING THE CONSERVING REGULATOR NOTE: For information on how to operate the cylinder, consult your equipment provider. NOTE: For this procedure, refer to FIGURE 3.1. 1. Loosen the cylinder post valve T‐Handle. 2. Lower the conserver over any CGA870 post valve cylinder with the alignment pins toward the holes on the cylinder neck. 3. Align the two pins and gasket with the corresponding holes on FIGURE 3.1 Attaching the the cylinder post valve.
SECTION 4—OPERATING INSTRUCTIONS SECTION 4—OPERATING INSTRUCTIONS DANGER NEVER smoke in an area where oxygen is being administered. NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere. CAUTION DO NOT use pediatric, low flow nasal cannulas or oxygen masks with the oxygen conserving regulator. DO NOT block the cannula connection or kink cannula tubing when the oxygen conserving regulator is in use. This may damage the oxygen conserving regulator. 1.
SECTION 4—OPERATING INSTRUCTIONS 4. Listen for leaks. If a leak is present, close the cylinder valve, check the CGA870 seal washer, and reinstall. If the leak persists, DO NOT use the equipment. Contact your Home Care Provider for repair. 5. Check the oxygen pressure gauge to verify the cylinder pressure. 6. Ensure that the vent hole on the side of the unit is not obstructed. 7. Position the cylinder so the oxygen cylinder valve outlet is pointing away from the user and any other person(s). 8.
SECTION 4—OPERATING INSTRUCTIONS WARNING NO OXYGEN is delivered in between settings. To obtain desired flow, the indicating pointer MUST align with a specific number on the dial. To avoid injury to patient, ALWAYS confirm prescribed setting before administering to patient and monitor flow on a frequent basis. 10. Turn the selector switch until indicating pointer is aligned with the prescribed setting. Refer to FIGURE 4.2 on page 17.
SECTION 4—OPERATING INSTRUCTIONS 12. When finished using the system, close the oxygen supply cylinder valve and continue breathing through the nasal cannula until no further oxygen flow is detected. 13. Remove the nasal cannula and turn the selector switch to the “0” OFF position. 14. When not in use, store in a clean, dry location. Continuous Flow Setting (CF) Selector Switch FIGURE 4.
SECTION 5—REMOVING THE CONSERVING REGULATOR SECTION 5—REMOVING THE CONSERVING REGULATOR DANGER NEVER attempt to remove the oxygen conserving regulator from an oxygen cylinder unless the cylinder valve is completely closed. CAUTION DO NOT try to loosen the T‐handle beyond the stop. Otherwise, damage to the oxygen conserving regulator will occur. NOTE: For information on how to operate the cylinder, consult your equipment provider. 1. Completely close the oxygen cylinder valve. 2.
SECTION 6—TYPICAL PRODUCT PARAMETERS SECTION 6—TYPICAL PRODUCT PARAMETERS OXYGEN DELIVERY: CONTINUOUS FLOW EMERGENCY BYPASS SETTING: SELECTOR SWITCH POSITION LITER FLOW EQUIVALENCY 0 OFF 1 1 2 2 3 3 4 4 5 5 6 6 Factory Preset at 2 ± .5 L/min. REQUIRED OPERATING PRESSURE: 200 psi to 3000 psi (13.8 bar to 206.8 bar) REGULATOR: Built-in, 25 ± 5 psi (1.7 bar ± .34 bar), brass high pressure with aluminum low pressure materials DIMENSIONS: Weight Approximately 14.
SECTION 6—TYPICAL PRODUCT PARAMETERS SHOCK VIBRATION: Not to exceed IEC 601-1 requirements Not to exceed IEC 68-2-6, IEC 68-2-34 No smoking or open flames Consult accompanying documents OXYGEN CYLINDER DURATION The total delivery of oxygen via the conserver is related to the breathing rates, it is user adaptive in that the total oxygen delivered per minute will automatically adjust with user need, as expressed by increased or decreased breathing rates.
SECTION 7—MAINTENANCE SECTION 7—MAINTENANCE CAUTION DO NOT use cleaning solutions. DO NOT immerse product in any kind of liquid. All repairs MUST be done by Invacare Corporation. 1. Disconnect all connections before cleaning. 2. After each use, clean exterior of the product with a clean, lint‐free cloth. 3. Store product in a clean area free from grease, oil, and other sources of contamination.
SECTION 8—TROUBLESHOOTING SECTION 8—TROUBLESHOOTING If oxygen conserving regulator fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider. PROBLEM No flow. Unit does not pulse. Unit flows or pulses continuously. POSSIBLE CAUSE SOLUTION 1. Cylinder valve is closed. 1. Turn the cylinder valve to the “ON” position. 2. Regulator in “Off” position. 2. Set to prescribed setting. 3. Cylinder empty. 3. Replace cylinder. 4.
LIMITED WARRANTY LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser/user of our products. This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state. Invacare warrants this product to be free from defects in materials and workmanship for a period of two years from date of purchase.
THAN INVACARE OR AN INVACARE SERVICE CENTER. SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE. THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN.