SECTION 7—CYLINDER FILL TIMES
HomeFill
™
22 Part No 1112648
SECTION 7—CYLINDER FILL TIMES
NOTE:Allfillingtimesareapproximateandmayvaryon
environmentalconditions.
*
Troubleshooting
IfConservingCylinderfailstofunction,consultthe
Troubleshootingguide.Ifproblemcannotbecorrected,
consultyourDealer/HomeCareequipmentprovider.
Replacement Parts - Accessories
HF2PCL6CarryingBag,HF2PC9CarryingBag,orHF2PCL4
CarryingBag
Cylinder
Filling
Times
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
L/min)
HF2PC9 HF2PCL4 HF2PCL6
NORMAL 2 hr 20 min 60 min 1 hr 25 min
MAXIMUM 2 hr 45 min 1 hr 20 min 1 hr 50 min
PROBLEM POSSIBLE CAUSE SOLUTION
No flow. 1. Regulator in “Off”
position.
1. Set to prescribed
setting.
2. Cylinder empty. 2. Replace/fill cylinder.
3. Dial not set at incre-
ment.
3. Set dial so indictor
points to a setting.
4. Conserver not sens-
ing breath.
4. Check position of
cannula in nose. Do not
breathe through mouth.
5. Check to see if the
cannula is kinked.