Product Manual
SECTION 5—OPERATING INSTRUCTIONS
Part No 1112648 19 HomeFill
™
SECTION 5—OPERATING
INSTRUCTIONS
WARNING
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
NOTE:Forthisprocedure,refertoFIGURE5.1.
1. Priortoeachuse,inspectproductforvisibledamage.DO
NOTuseifanydamageisfound.
NOTE:Ifforanyreason,anylabelbecomesillegibleorlost,contact
yourdealer/HomeCareequipmentprovider.
2. Besureregulatorisinthe“Off”position.Align“Off”with
theindicatingpointer.
3. Positionthecylindersotheoxygencylindervalveoutletis
pointingawayfromtheuserandanyotherperson(s).
WARNING
DONOTdirectflowofoxygenatanyperson,orflammable
materialwhenadjustingtheflowdial.
4. Attachastandardadultsinglelumenoxygennasal
cannula,nolongerthansevenfeettotheproduct’soutlet
accordingtothecannulamanufacturer’sinstructions.DO
NOTusepediatric,lowflownasalcannulasoroxygen
maskswiththisproduct.
NOTE:Aunitwithoutcannulaattachedmaypulseifsetatposition
1‐5.