Product Manual
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
™
14 Part No 1112648
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
WARNING
NEVERusetoolsofanykindtoconnect/disconnectthe
cylinderandthecompressor.Otherwise,severeinjuryand/or
damagemayoccur.
DONOTdropoxygencylinders.Usetwohandswhen
handling/transportingoxygencylinders.Otherwise,injuryor
damagemayoccur.
DONOTusealiquidleakdetectortotestforleaks.
NOTE:Forthisprocedure,refertoFIGURE4.2andFIGURE4.3.
1. Examinethecylindergauge.Ifthecylinderpressureisless
than1500psig.,proceedtoSTEP2(FIGURE4.3).
NOTE:Ifthecylinderpressureisgreaterthan1500psig.,DO
NOTattempttotopoffthiscylinder.Itmaynotfill.
2. SettheregulatorflowdialonthecylindertoOFF(FIGURE
4.3).
3. Removethecylinderandconnectorfillportcovers(if
present).
CAUTION
DONOTconnecttheoutlet(cannula)connectiontothe
HomeFillunit,otherwisetheconservingcylinderwillnot
workproperly.
4. MomentarilypushDOWNontheouterring(sleeve)ofthe
connectorfillporttoresettheconnectorasshownin
FIGURE4.2.