Operator’s Manual HomeFill™ Oxygen Conserving Cylinder HF2PCL6 HF2PC9 HF2PCL4 Dealer: This manual MUST be given to the end user. User: Before using this cylinder, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING WARNING DO NOT OPERATE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT.
TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES ..........................................................4 RECEIVING INSPECTION .............................................5 INTENDED USE ...........................................................5 SECTION 1—GENERAL GUIDELINES .............................6 Operating Information.............................................................................................6 Maintenance .......................................................................
SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
RECEIVING INSPECTION CAUTION “Caution: Federal law restricts this device to sale by or rental or on the order of a physician, or any other practitioner licensed by the law of the state in which he/she practices to use or order the use of this device.” RECEIVING INSPECTION Remove product from package and inspect for damage. If there is any damage, DO NOT use. Contact your dealer/ Home Care equipment provider.
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the HomeFill Oxygen Conserving Cylinder these instructions MUST be followed. WARNING This section contains important information for the safe operation and use of this product. Operating Information Only personnel instructed and trained in the use of the HomeFill O2 Conserving Cylinder should operate this product.
SECTION 1—GENERAL GUIDELINES NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere. ALWAYS follow ANSI and CGA standards for Medical Gas Products and Flowmeters (E‐7) and Oxygen Handling (G‐4). NO OXYGEN is delivered when the pointer is aligned with “Off”. DO NOT block the outlet fitting or kink the cannula when the product is in use, this may damage the unit. The use of this device is limited to the oxygen patient.
SECTION 1—GENERAL GUIDELINES DO NOT use a humidifier with this product. DO NOT allow cylinders to tip or fall. Secure gas cylinders so they cannot fall. Ensure cylinder is always in upright position. DO NOT store cylinders near sources of heat or flame. This product is equipped with a relief valve. If you hear a loud hissing or popping sound coming from the product, discontinue use of product, turn the flow selector to the “Off” position, and contact your dealer/Home Care equipment provider.
SECTION 2—FEATURES SECTION 2—FEATURES DANGER This product is not intended as a life‐sustaining or life‐supporting device. Outlet Connection Cannula Flow Dial Cylinder Fill Port Cylinder Gauge FIGURE 2.
SECTION 3—TYPICAL PRODUCT PARAMETERS SECTION 3—TYPICAL PRODUCT PARAMETERS CONSERVER S/N RANGE 1-69,999 75,000 + INLET PRESSURE RANGE: 300 - 2100 psig 300 - 2100 psig 50 psig 50 psig HF2PCL6: Weight 4.3 lbs (1.95 kg) 4.1 lbs (1.86 kg) HF2PCL6: Overall Length 11.1 in (28 cm) 10.9 in (27.7 cm) HF2PCL6: Diameter 4.4 in (11.1 cm) 4.4 in (11.1 cm) HF2PC9: Weight 5.3 lbs (2.4 kg) 5.1 lbs (2.31 kg) HF2PC9: Overall Length 14.2 in (36 cm) 14 in (35.5 cm) HF2PC9: Diameter 4.4 in (11.1 cm) 4.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: Refer to the HomeFill Compressor Owner’s Manual for compressor operating instructions. Cylinder Prefill Inspection WARNING All cylinders MUST be inspected before attempting to fill; otherwise, injury or damage may occur. NEVER smoke in an area where oxygen is being administered. NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE External Examination 1. Examine the outside of the cylinder for the following conditions, and replace the cylinder if they exist: • Dents or dings • Arc Burns • Oil or Grease • Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use. 2. Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or blistering of the paint, or other protective coating or heat sensitive indicator.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Improper filling or abuse of this cylinder, or failure to follow these safety instructions may cause serious bodily injury or death. High pressure oxidizing gas vigorously accelerates combustion. DO NOT alter or modify cylinder or related components. DO NOT use oil or grease. Keep cylinder, cylinder valve, all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death. For medical applications.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Connecting/Disconnecting the Cylinder to/from the Compressor Connecting the Cylinder to the Compressor WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. DO NOT drop oxygen cylinders. Use two hands when handling/transporting oxygen cylinders. Otherwise, injury or damage may occur. DO NOT use a liquid leak detector to test for leaks.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport. PUSH DOWN Outer Ring (Sleeve) Connector Fillport GREEN Dots BEFORE COUPLING CYLINDER, PUSH DOWN ON SLEEVE UNTIL GREEN DOTS ARE VISIBLE. FIGURE 4.2 Resetting Connector Fillport 5.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Compressor Cylinder/Regulator Assembly Cylinder Fillport Connector Fillport Outer Ring (Sleeve) Cylinder Gauge PULL UP Regulator Cylinder Cylinder Gauge Regulator Flow Dial Outlet (Cannula) Connection Cylinder Fillport Power Switch Compressor Cradle Outer Ring (Sleeve) Connector Fillport NOTE: Fillport covers not shown for clarity. FIGURE 4.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Disconnecting the Cylinder from the Compressor WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. NOTE: For this procedure, refer to FIGURE 4.4. 1. Press the compressor power switch to the Off (O) position. 2. Grasp the cylinder/regulator assembly in the area behind the cylinder gauge. 3.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE 5. When the cylinder fillport is disconnected from the connector fillport, release the outer ring (sleeve) of the connector fillport and use two hands to remove the cylinder from the compressor cradle. 6. Place the fillport covers onto the connector and cylinder fillports. NOTE: The fillport covers should be used whenever cylinders are not being filled.
SECTION 5—OPERATING INSTRUCTIONS SECTION 5—OPERATING INSTRUCTIONS WARNING NEVER smoke in an area where oxygen is being administered. NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere. NOTE: For this procedure, refer to FIGURE 5.1. 1. Prior to each use, inspect product for visible damage. DO NOT use if any damage is found. NOTE: If for any reason, any label becomes illegible or lost, contact your dealer/Home Care equipment provider. 2.
SECTION 5—OPERATING INSTRUCTIONS 5. Place the cannula over your ears and position the prongs in your nose as instructed by your Health Care provider or cannula manufacturer. Refer to FIGURE 5.1. 6. Turn dial until indicating pointer is aligned with the prescribed setting. FIGURE 5.1 Cannula Placement WARNING When pointer on contents gauge enters red zone, it is recommended to refill the cylinder on the Invacare HomeFill Compressor. NO OXYGEN is delivered in between settings.
SECTION 6—MAINTENANCE SECTION 6—MAINTENANCE 1. After each use, clean exterior of the product with a dry, lint free cloth only. 2. Store product in a clean area free from grease, oil, and other sources of contamination. CAUTION DO NOT use cleaning solutions. DO NOT immerse product in any kind of liquid. All repairs MUST be done by Invacare Corporation.
SECTION 7—CYLINDER FILL TIMES SECTION 7—CYLINDER FILL TIMES NOTE: All filling times are approximate and may vary on environmental conditions. * Cylinder Filling Times Concentrator Flow Rate to Patient: (Platinum 5 up to 3 L/min. and Platinum 10 up to 5 L/min) HF2PC9 HF2PCL4 HF2PCL6 NORMAL 2 hr 20 min 60 min 1 hr 25 min MAXIMUM 2 hr 45 min 1 hr 20 min 1 hr 50 min Troubleshooting If Conserving Cylinder fails to function, consult the Troubleshooting guide.
LIMITED WARRANTY LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser/user of our products. This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state. Invacare warrants this product to be free from defects in materials and workmanship for a period of three years from date of purchase.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN.