Product Manual

10 TROUBLESHOOTING
Part No 1123842 51 Invacare
®
IVC
Model BAR600IVC Bariatric Bed
10.2 Troubleshooting: Mechanical
SYMPTOM PROBABLE CAUSE SOLUTIONS
Bed does not stay in place. Locking casters not locked. Lock casters. Refer to Installing Casters on page 35.
Bed sections do not move. Pull tube Ensure pull tubes are properly connected. Refer to FIGURE11.
Check pull tube(s) for binding. Call dealer for repairs.
Bed ends move in opposite
directions
(electrical or manual bed).
Drive shaft connected to wrong gear
box shaft.
Reinstall drive shaft
. Refer to Assembling/Installing/Removing
Drive Shaft
on page 40
.
Hi/Lo function does not work. Drive shaft disconnected.
Reinstall drive shaft.
Refer to Assembling/Installing/Removing
Drive Shaft
on page 40
.
Pressing function on pendant results in
motion other than that intended.
Motor cable connections to junction
box incorrect.
Verify connections. Refer to FIGURE11.