8 PACKAGING, HANDLING AND SET-UP
Part No 1123842 35 Invacare
®
IVC
Model BAR600IVC Bariatric Bed
Installing Casters
1. To install casters, lay the bed ends flat on the floor.
2. Insert the caster shaft into the caster socket.
3. To prevent excess movement of the bed during assembly, lock each of the two locking casters by pushing down on the caster lock.
4. To unlock each of the two locking casters, push down on the caster lock labeled off.
WARNING
Ensure that the locking casters are installed diagonally from each other and are locked. Otherwise, damage or personal
injury could occur.
The proper casters are provided with the Bariatric Bed. DO NOT use any other casters with the Bariatric Bed. Otherwise,
damage or personal injury could occur.
Ensure that casters are locked and that the bed does not move before transferring. DO NOT transfer occupant if bed is
not stable.
For this procedure, refer to FIGURE6 on page 36.
ALWAYS locate locking caster in the same location on each bed end so that they will always be assembled diagonally
opposite each other when inventory is mixed.
Position the two locking casters so that they will be diagonally opposite from one another when the bed is assembled.
If two locking casters are not included with the bed, contact an Invacare dealer.
Locking the casters may not prevent the bed from moving on slick or slippery surfaces.