8 PACKAGING, HANDLING AND SET-UP
Part No 1123842 31 Invacare
®
IVC
Model BAR600IVC Bariatric Bed
FIGURE3 Remove Tie Wraps
4. Align the inside of the head section with the inside of the foot section.
5. Push the head section towards the foot section so that the head section frame slides into the foot section frame and the locking pins engage.
6. Ensure the locking pins are properly engaged by attempting to pull the head section from the foot section. If the sections separate, repeat STEP 5-6
until the locking pins are secure.
7. Install the bed ends. Refer to Installing/Removing Bed Ends
on page 34.
8. Connect the crossover cable. Refer to Connecting Crossover Cable
on page 37.
9. Assemble and install the drive shaft. Refer to Assembling/Installing/Removing Drive Shaft
on page 40.
10. Activate the bed functions. Refer to Inspecting the Bed on page 44.
Locking pins are located on the foot section frame and will make an audible click when properly engaged with the head
section frame.
Tie Wrap
Tie Wrap