Product Manual
6 TROUBLESHOOTING
Invacare
®
IVC
™
Model BAR600IVC Bariatric Bed 24 Part No 1123842
6 Troubleshooting
6.1 Troubleshooting: Electrical
WARNING
Read all general guidelines, warnings and cautions before performing any repair/maintenance on the bed. All repair and
maintenance on the bed should be performed by a qualified technician.
SYMPTOM PROBABLE CAUSE SOLUTIONS
Bed section does not move. End of stroke reached. Operate opposite button.
Full Electric Bed: Hi/Lo func-
tion does not work.
Bed not plugged in. Ensure power cord is plugged into power source.
Pendant not functioning. Ensure Pendant correctly connected to
junction box.
Ensure Pendant cable is not entangled/pinched.
Cables Ensure cables are not entangled/pinched.
Call the dealer for repair/replacement of cables and/or motors.
Household fuse or breaker blown. Check household fuse/breaker box.
Motors are not functioning. Call dealer.
Hi/Lo Motor Ensure cables are not entangled or pinched.
Ensure drive shaft is correctly installed.
Bed Spring does not stop
moving.
Pendant Check for depressed/stuck buttons. If button is stuck, call dealer for
repair.
Bed Hi/Lo movement does
not stop.
Check cables. If damaged, call dealer.