5 MAINTENANCE
Invacare
®
IVC
Model BAR600IVC Bariatric Bed 22 Part No 1123842
5.2 Cleaning Head and Foot Sections
1. Unplug the power cord of electric beds before performing maintenance.
2. Wipe down the bed with disinfectant. Allow it to dry completely.
5.3 Maintenance and Cleaning Checklist
Invacare recommends the following maintenance and cleaning procedures be conducted between uses and at least once every three years. This form
is provided as a guide to help with documentation.
Cleaning and Disinfecting Bed
Unplug the power cord of electric beds before performing maintenance.
Wipe down with disinfectant. Allow to dry completely.
Mechanical Inspection and Maintenance
Inspect all bed components for damage or excessive wear.
Visually examine all welds for cracks.
Inspect the head and foot spring sections for bending, warping or damage.
Check drive shaft and drive shaft connections for bending, damage or excessive wear.
Inspect pull tubes and mounting hardware for bending, damage or excessive wear.
Inspect all bolts and rivets to ensure that they are securely tightened and functioning properly.
Check sleep surfaces to ensure all links are intact.
Check casters to ensure they lock, if applicable, and roll properly.
Electrical Inspection and Maintenance
Inspect all electrical bed components for damage or excessive wear (i.e., cracked or broken housings, or worn components).
Check pendant, power and motor cords for chafing, cuts or excessive wear.
Make sure all plugs are fully attached and free of damage.