User's Manual
Chapter 3 Performing Capsule Endoscopy
3-15
After washing hands, patients should grip the gold section of the
MiroCam® Capsule between the thumb and the forefinger of the
both hands.
CAUTION
Initialize the MiroCam® Receiver prior to the MiroCam®
Capsule Endoscopy Procedure. Once the MiroCam®
Receiver confirms signal transmissions with a green SIG
indicator, the receiver will start recording for 12 hours, even
while the power switch of the MiroCam® Receiver is in the
off position.
3.5.7 Checking Operation of the MiroCam® Receiver and Signal
Reception
After the patient holds the MiroCam® Capsule as detailed above,
the signal indicator on the MiroCam® Receiver is green and signal
indication melody is generated. If signal indicator is yellow, there is
no signal being received from the capsule.
Switch on the MiroCam® Receiver.
Check if the battery indicator (BAT) at the top of the MiroCam®
Receiver shows that the battery is fully charged. The battery
indicator on the MiroCam® Receiver should be green.
Check if the initialize indicator (INI) at the top of the MiroCam®
Receiver shows that initialization is complete. The INI indicator on
the MiroCam® Receiver should be green.