User's Manual
Chapter 3 Performing Capsule Endoscopy
3-5
3.3.2 Initialize the MiroCam® Receiver
Connect the smaller side of the USB Cable to the MiroCam®
Receiver
Connect the larger side of the USB cable to the workstation.
Turn on the MiroCam® Receiver.
Click the
in the 0LUR9LHZ Operator to initialize the
MiroCam® Receiver.
Click the
in the 0LUR9LHZ Operator to input patient
information.
After completing the required fields, click the
to save the
patient information onto the MiroCam® Receiver.
WARNING
If the MiroCam® Receiver is initialized, any patient data
stored on the device will be deleted permanently.
NOTE
During initialization with MR1100, it is not possible to cancel
the process.
3.3.3 MiroCam® Receiver Preparation
Connect the battery to the MiroCam® Receiver.
- Insert pins and bumps on back side of the MiroCam®
Receiver in the slots on the battery.
- Press battery to the MiroCam® Receiver until clasp locks
battery in place.
Turn on the power switch of the MiroCam® Receiver.
Check the battery indicator (BAT) of upper side of the MiroCam®
Receiver. If the battery is fully charged, the LED light will be
green. If the LED light is not green, charge the battery.