3.5.6 Checking Operation of the Receiver and Signal Reception
After the patient holds the capsule as detailed above, the signal
indicator on the receiver is green and signal indication melody is
generated.. If signal indicator is yellow, there is no signal being
received from the capsule.
n Switch on the receiver.
n Check if the battery indicator (BAT) at the top of the receiver shows
that the battery is fully charged. The battery indicator on the
receiver should be green.
n Check if the initialize indicator (INI) at the top of the receiver shows
that initialization is complete. The INI indicator on the receiver
should be green.
n Connect the data cable to the MiroCam® receiver.
n After patient grasps capsule by opposing gold bands (as detailed
in picture above, check the SIG indicator on the receiver is green
and signal indication melody is generate.
CAUTION
If the battery indicator on the receiver shows that the
battery is not fully charged, do not start the endoscopy
procedure. The battery may be depleted during the
procedure and images may not be saved properly.
CAUTION
Please be noticed to confirm the shape and the direction of
the connector when you connect the data cable to
MiroCam® Receiver, since it may cause cable damage
problem.