Chapter 3 Performing Capsule Endoscopy
3- 12
n Make sure that the capsule packaging is not damaged in any way.
When removed capsule from case, the capsule should be flashing.
CAUTION
Do not use a capsule endoscope when the packaging or
the capsule is damaged.
n Check if the front of the capsule is flashing when it is removed
from the case
n After washing hands, patients should grip the gold section of the
capsule between the thumb and the forefinger of the both hands.
NOTE
After turning the receiver power on, if a signal is not received
by the receiver within 50 seconds (i.e. the SIG indicator does
not turn green), the SIG indicator will flash orange and the
receiver will make a ‘beep’ ‘beep’ warning sound. This will
occur until a signal is received or the power is turned off.
In this case, you must begin the procedure immediately or