User's Manual

Chapter 3 Performing Capsule Endoscopy
3- 16
3.7 Receiver Management – Post Procedure
3.7.1 Cleaning
Remove data cable from receiver unit.
Capsule and sensor pads should not be reused.
To clean the MiroCam® Receiver unit and Workstation, dampen a
soft cloth with warm water or with a commercial, nonabrasive
cleaner (or a mild soap or detergent solution) and wipe the exterior
surface lightly. Do not allow any liquids to come in contact with the
power connector, cable connector or switches. Do not allow any
liquids to penetrate connectors or openings in the system cover.
Cleaning of the system and accessories should be conducted at
least once a week.
CAUTION
Do not immerse the MiroCam® capsule endoscope system
or its accessories in liquid or clean with caustic or abrasive
cleaners. Do not spray or pour any liquid on the system or
its accessories. Do not use the organic solvents such as a
lacquer, thinner, ethylene and oxide because they can
damage the equipment. When cleaning the data cables, do
not use the cleaning tools such as brush and sandpaper.
3.7.2 Charging Battery
Battery must be recharged after every procedure.
Battery can be individually recharged or with receiver unit.
Battery must be recharged in battery charger.
If two batteries are inserted in charger, the batteries will
automatically consecutively charge (not simultaneous).