User's Manual
Performing Capsule Endoscopy Chapter 3
3-5
Note
Before clicking the
icon in MiroView™, connect USB
cable to MiroCam® Receiver unit.
WARNING
If MiroCam® receiver unit is initialized, any patient data
stored on the device will be deleted permanently.
3.3.3 Receiver Preparation
Connect battery to the receiver unit.
- Insert pins on back side of the receiver unit in the slots on the
battery.
- Press battery to receiver until clasp locks battery in place.
Turn on power switch of receiver unit.
Check battery indicator (BAT) of upper side of receiver unit. If
battery is fully charged, light will be green. If light is not green,
charge the battery.
Check initialization indicator (INI) of upper side of receiver unit.
Prior to beginning a procedure, this light must be green.
Connect data cable to receiver unit.
Indicator Description
SIG
Indicates status of signal from capsule
Green : Signal is being received from capsule
Yellow : Signal is not being received from capsule
INI
Initialization status of Receiver Unit
Green : Receiver is initialized
Yellow : Receiver is not initialized
BAT
Battery Status
Green : Fully charged
Yellow : Not charged
WARNING
DO NOT do begin procedure when receiver unit’s battery
indicator is not green. This indicates battery is not fully
charge or out of order, and patient data may be lost.