System Handbook

~ "
~
TTL
In
Collaboration
with
s p e a g
CALIBRA
TI
ON
LABORATORY
Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
Tel: +86-10-62304633-2079 Fax: +86-10-62304633-2504
E-mail: cttl@chinattl.com http :
//
www
. chinattl. en
DASY5 Validation Report for Body TSL Date: 03.07.2019
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 - SN: 1107
Communication System: UID 0, CW; Frequency: 750 MHz; Duty Cycle:
1:
1
Medium parameters used:
f=
750 MHz;
cr
= 0.943 S/m; Er= 54.78; p = 1000 kg/
m3
Phantom section: Center Section
DASY5 Configuration:
Probe: EX3DV4 - SN3617; ConvF(9.85, 9.85,
9.85)@
750 MHz; Calibrated:
1/31/2019
Sensor-Surface: 1.4mm (Mechanical Surface Detection)
Electronics: DAE4 Sn 1331; Calibrated: 2/6/2019
Phantom: MFP _
V5.1C;
Type: QD 000 P51CA; Serial: 1062
Measurement SW: DASY52, Version 52.
10
(2); SEMCAD X Version 14.6.12
(7450)
Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube
0:
Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 52.31 V /m; Power
Drift=
0.00 dB
Peak SAR (extrapolated)= 3.09 W/kg
SAR(l
g) = 2.09 W/kg; SAR(lO g) = 1.4 W/kg
Maximum value
of
SAR (measured)= 2.
75
W/kg
dB
0
-1.97
-3.93
-5.90
-7.86
-9.83
0
dB=
2.75 W/kg = 4.39 dBW/kg
Certificate No: 219-60054
Page
7
of8