DASY5 Validation Report for Head TSL
Date: 25.11.2019
Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 - SN :4d167
Communication System: UID O - CW; Frequency: 835 MHz
Medium parameters used: f = 835 MHz;
cr
= 0.91 Sim;
tr=
42; p = 1000 kg/m
3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)
DASY52 Configuration:
• Probe: EX3DV4 - SN7349; ConvF(9.89, 9.89, 9.89)
@
835
MHz; Calibrated: 29.05.2019
• Sensor-Surface: 1.4mm (Mechanical Surface Detection)
• Electronics: DAE4 Sn601; Calibrated: 30.04.2019
• Phantom: Flat Phantom 4.9 (front); Type: QD
OOL
P49 AA; Serial: 1001
• DASY52 52.10.3(1513); SEMCAD X 14.6.13(7474)
Dipole Calibration for Head Tissue/Pin=250 mW, d=l5mm/Zoom Scan (7x7x7)/Cube
0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference
Value=
63.15 V/m; Power
Drift=
-0.00
dB
Peak SAR (extrapolated)= 3.59 W/kg
SAR(l
g) = 2.4 W/kg; SAR(lO g) = 1.56 W/kg
Smallest distance from peaks to all points 3
dB
below =
16
mm
Ratio
of
SAR at M2 to SAR at
Ml=
66.8%
Maximum value
of
SAR (measured) = 3.20 W /kg
dB
0
-2.00
-4.00
-6.00
-8.00
-10.00
0
dB=
3.20 W/kg = 5.05 dBW/kg
Certificate No: D835V2-4d167 _Nov19 Page 5 of 7