Operator's Manual

Table Of Contents
Preparing the Monitor
78
BIS™ Advance Monitor
Number Description
1 Sensor
2 PIC
3 BISx module
4 MIC
5 Adapter cable
6 Monitor
7 Docking station
8
GCX mounting accessory (the selected accessory
will depend on the use case)
3.6.2. Connecting a Sensor to a
Patient
Select the appropriate sensor. A list of sensors appears in Table 34. BIS™ Advance
Accessories on page 252. It is important to use the correct sensor for your current
patient and monitoring situation, as follows:
Table 4. Sensor/BISx/Use Case Matrix
Sensor BISx module Monitoring Type
Quatro, Extend or
Pediatric sensor
BISx module or a BISx4
module
Two-channel monitoring
(in one hemisphere of
the brain)
Bilateral sensor BISx4 module
Four-channel monitoring
(in both hemispheres of
the brain).