Operator's Manual

Table Of Contents
Mounting the Docking Station
BIS™ Advance Monitor
GCX desktop mount) before or after you place the monitor in the docking
station.
2. GCX clamp mount or GCX desktop mount with required screws (screws are
packaged with the mount)
3. If desired, IV pole, anesthesia mount, or other hardware for mounting
medical devices for use
To set up the monitor with the GCX clamp mount:
1. Place the docking station on a stable surface.
2. If your docking station has four round black supports attached to its back
(to the four screw holes seen in Figure 15. Setup with GCX Clamp Mount on
page 66), remove these supports and set them aside in case you need
them in the future for other mounting solutions.
3. Locate the four screws supplied with the GCX clamp mount. Using a
standard screwdriver, attach the GCX clamp mount to the back of the
docking station using these four screws, by inserting the screws through
the four holes in the GCX clamp mount and into the four screw holes on
the back of the docking station (from which the four supports were
removed). Reference Figure 15. Setup with GCX Clamp Mount, below.
4. Once the GCX clamp mount and the docking station are firmly attached,
mount the GCX clamp mount to an IV pole, anesthesia mount or other
hardware, following the instructions supplied with that hardware.
5. The monitor can now be placed in the docking station and attached to a
BISx module and sensor to begin monitoring.