Operator's Manual
Table Of Contents
- 1. Introduction
- 1.1. Introduction
- 1.2. Safety Information
- 1.3. Device Description
- 1.4. Intended Purpose
- 1.5. BIS™ Advance Monitor System Indications and Contraindications
- 1.6. Patient Target Groups
- 1.7. Intended Users
- 1.8. Essential Performance
- 1.9. Adverse Events (Residual Risks)
- 1.10. The BIS™ Advance Parameters
- 1.11. Symbols
- 2. The BIS™ Advance Monitor — Equipment and Supplies
- 2.1. The BIS™ Advance System
- 2.2. Parts of the System
- 3. Installation and Preparation for Use
- 3.1. Operating Environment
- 3.2. Power Requirements and Battery Use
- 3.3. Battery Operation
- 3.4. Mounting the Docking Station
- 3.5. Preparing the Docking Station
- 3.6. Preparing the Monitor
- 3.7. Home Screen
- 3.8. Four-Channel Monitoring
- 3.9. Home Screen Trend Graphs
- 3.10. Home Screen Numeric Section
- 3.11. Home Screen Menus
- 3.12. Home Screen Options
- 3.13. Secondary Trend Selection
- 3.14. Home Screen Audio Alarm Settings
- 3.15. Settings and Maintenance
- 3.16. Info Options
- 3.17. Quick Reference Checklist
- 4. Operating the BIS™ Advance Monitoring System
- 4.1. Preparing for Operation
- 4.2. Cases
- 4.3. Alarms and Messages
- 4.4. Chart Data
- 4.5. Menu Map
- 4.6. Institutional Settings
- 5. Data Storage, Transfer, and Export
- 5.1. Cybersecurity and Data Integrity
- 5.2. Monitor Data Memory
- 5.3. BISx Data Memory
- 5.4. Types of Exported Data
- 5.5. Live Case Export
- 5.6. Saved Cases Export
- 5.7. Recording a Snapshot Event
- 5.8. BIS™ Saved Data Files
- 5.9. Logs Export
- 5.10. Viewing and Printing Saved Data in PDF Format
- 6. Service and Maintenance
- 6.1. Introduction
- 6.2. Servicing the BIS™ Advance Monitor
- 6.3. Battery Handling
- 6.4. Cleaning the BIS™ Advance Monitor
- 6.5. Instrument Identification
- 7. Diagnostics and Troubleshooting
- 7.1. Diagnostics
- 7.2. Troubleshooting
- 7.3. Administrator Mode
- 7.4. Demo Mode
- 7.5. Glossary
- 7.6. Specifications
- 7.6.1. General Specifications
- 7.6.2. Operating Environments
- 7.6.3. Power Supply
- 7.6.4. Battery Specifications
- 7.6.5. Controls and Connectors
- 7.6.6. Alarms
- 7.6.7. Display
- 7.6.8. Sound Pressure Data
- 7.6.9. BIS Specifications
- 7.6.10. EEG Specifications
- 7.6.11. BISx Module Specifications
- 7.6.12. Software/GUI Specifications
- 7.7. Electromagnetic Compatibility Specifications
- 7.8. Product Compliance
- 7.9. BIS™ Advance Components, Accessories, and Documentation
- 7.10. Warranty
- 7.11. Software License Agreement
Essential Performance
BIS™ Advance Monitor
27
For more information and technical support, contact
BISTechnicalsupport@medtronic.com.
1.8. Essential Performance
The essential performance of the monitor is the display of the BIS™ index and EEG in
order to monitor a patient's level of consciousness.
In addition, the monitor meets the essential performance requirements per IEC
80601-2-26:2019, as listed below.
Requirement Subclause Results
Accuracy of signal
reproduction
201.12.1.102 2-channel max variation of the output
signal is -13% of the input signal
4-channel max variation of the output
signal is -6.5% of the input signal
Input dynamic range
and differential offset
voltage
201.12.1.103 2-channel max deviation of the output
signal with applied +300V DC (DC offset
voltage) is 1.6% compared to the output
signal without applied DC offset voltage
2-channel max deviation of the output
signal with applied -300V DC (DC offset
voltage) is 3.4% compared to the output
signal without applied DC offset voltage
4-channel max deviation of the output
signal with applied +300V DC (DC offset
voltage) is 1.2% compared to the output
signal without applied DC offset voltage
4-channel max deviation of the output
signal with applied -300V DC (DC offset
voltage) is 0.4% compared to the output
signal without applied DC offset voltage
Input noise 201.12.1.104
4-channel: The max measured signal
noise is: 3.5μV (peak-to-valley)
2-channel: The max measured signal
noise is: 1.3μV (peak-to-valley)