Operator's Manual

Table Of Contents
Electromagnetic Compatibility Specifications
242
BIS™ Advance Monitor
The BIS™ Advance monitor is intended for use in the electromagnetic
environment specified below. The customer or user of the BIS™ Advance
monitor should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
Note: U
T
is the AC mains voltage prior to the application of the test level.
Conducted RF
IEC 61000-4-6
3 Vrms, 150
kHz to 80 MHz
6 Vrms (ISM
Bands)
3 Vrms, 150
kHz to 80 MHz
6 Vrms (ISM
Bands)
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the BIS™ Advance
monitor, including cables,
than the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
d = 1.2 √P
d = 1.2 √P,
80 MHz to 800 MHz
d = 2.3 √P,
800 MHz to 2.5 GHz
where P is the maximum
output power rating of
the transmitter in watts
according to the
transmitter manufacturer
and d is the
recommended separation
distance in meters (m).
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.7 GHz
3 V/m
80 MHz to
2.7 GHz