Operator's Manual

Table Of Contents
Electromagnetic Compatibility Specifications
BIS™ Advance Monitor
The BIS™ Advance monitor is intended for use in the electromagnetic
environment specified below. The customer or user of the BIS™ Advance
monitor should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
U
T
= 0%, 1
cycle
U
T
= 70%,
25/30 cycles (0
degrees)
U
T
= 0%, 1
cycle
U
T
= 70%,
25/30 cycles (0
degrees)
during power mains
interruptions longer than
45 minutes, it is
recommended that the
BIS™ Advance monitor be
powered by an
uninterruptible power
supply.
Short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
U
T
= 0%,
250/300 cycles
U
T
= 0%,
250/300 cycles
Mains power quality
should be that of a typical
hospital environment. If
the user of the BIS™
Advance monitor requires
continued operation
during power mains
interruptions longer than
45 minutes, it is
recommended that the
BIS™ Advance monitor be
powered by an
uninterruptible power
supply.
Power
frequency
(50/60 Hz)
magnetic field.
IEC 61000-4-8
30 A/m, 60 Hz 30 A/m, 60 Hz Power frequency
magnetic fields should be
at levels characteristic of
typical location in a
typical hospital
environment.