Operator's Manual
Table Of Contents
- 1. Introduction
- 1.1. Introduction
- 1.2. Safety Information
- 1.3. Device Description
- 1.4. Intended Purpose
- 1.5. BIS™ Advance Monitor System Indications and Contraindications
- 1.6. Patient Target Groups
- 1.7. Intended Users
- 1.8. Essential Performance
- 1.9. Adverse Events (Residual Risks)
- 1.10. The BIS™ Advance Parameters
- 1.11. Symbols
- 2. The BIS™ Advance Monitor — Equipment and Supplies
- 2.1. The BIS™ Advance System
- 2.2. Parts of the System
- 3. Installation and Preparation for Use
- 3.1. Operating Environment
- 3.2. Power Requirements and Battery Use
- 3.3. Battery Operation
- 3.4. Mounting the Docking Station
- 3.5. Preparing the Docking Station
- 3.6. Preparing the Monitor
- 3.7. Home Screen
- 3.8. Four-Channel Monitoring
- 3.9. Home Screen Trend Graphs
- 3.10. Home Screen Numeric Section
- 3.11. Home Screen Menus
- 3.12. Home Screen Options
- 3.13. Secondary Trend Selection
- 3.14. Home Screen Audio Alarm Settings
- 3.15. Settings and Maintenance
- 3.16. Info Options
- 3.17. Quick Reference Checklist
- 4. Operating the BIS™ Advance Monitoring System
- 4.1. Preparing for Operation
- 4.2. Cases
- 4.3. Alarms and Messages
- 4.4. Chart Data
- 4.5. Menu Map
- 4.6. Institutional Settings
- 5. Data Storage, Transfer, and Export
- 5.1. Cybersecurity and Data Integrity
- 5.2. Monitor Data Memory
- 5.3. BISx Data Memory
- 5.4. Types of Exported Data
- 5.5. Live Case Export
- 5.6. Saved Cases Export
- 5.7. Recording a Snapshot Event
- 5.8. BIS™ Saved Data Files
- 5.9. Logs Export
- 5.10. Viewing and Printing Saved Data in PDF Format
- 6. Service and Maintenance
- 6.1. Introduction
- 6.2. Servicing the BIS™ Advance Monitor
- 6.3. Battery Handling
- 6.4. Cleaning the BIS™ Advance Monitor
- 6.5. Instrument Identification
- 7. Diagnostics and Troubleshooting
- 7.1. Diagnostics
- 7.2. Troubleshooting
- 7.3. Administrator Mode
- 7.4. Demo Mode
- 7.5. Glossary
- 7.6. Specifications
- 7.6.1. General Specifications
- 7.6.2. Operating Environments
- 7.6.3. Power Supply
- 7.6.4. Battery Specifications
- 7.6.5. Controls and Connectors
- 7.6.6. Alarms
- 7.6.7. Display
- 7.6.8. Sound Pressure Data
- 7.6.9. BIS Specifications
- 7.6.10. EEG Specifications
- 7.6.11. BISx Module Specifications
- 7.6.12. Software/GUI Specifications
- 7.7. Electromagnetic Compatibility Specifications
- 7.8. Product Compliance
- 7.9. BIS™ Advance Components, Accessories, and Documentation
- 7.10. Warranty
- 7.11. Software License Agreement
Specifications
236
BIS™ Advance Monitor
Default 6.25 mm/sec
Computed Parameters
Bispectral Index, Suppression Ratio, Suppression
Time, EMG, Signal Quality Indicator, and Burst
Count
User-defined Displays Trend and real-time EEG waveforms
Update Rate 1 second for BIS™ number, 10 seconds for Trend
Alarms Auditory and visual, user adjustable limits
Filters ON (2-70 Hz with notch) or OFF (0.25-100 Hz)
Mode
Sensor Automatically selects mode two-channel
or 4-channel)
7.6.11. BISx Module Specifications
Item
Value
Weight 10.0 oz (0.284 kg) including integral cable
Dimensions 3.75 in. (9.5 cm) diameter x 2.5 in. (6.3 cm) thick
Cable Length
9 ft (2.7 m) Integral BISx Cable
4 ½ ft (1.4 m) from BISx to sensor connector
Analog to Digital Converter Noise-shaped sigma-delta
Sampling Rate 16,384 samples/second
Resolution 16 Bits at 256 samples/second
Input Impedance
50 M ohms typical (DC)
5 M ohms typical (at 10 Hz)
Noise < 0.3 μV RMS (2.0 μV peak-to-peak);