Operator's Manual

Table Of Contents
Specifications
228
BIS™ Advance Monitor
Item Value
800 mm Hg (-457 [1500ft] below sea level) to
360mm Hg (6100 [20,000ft] above sea level) for
BISx module
Operating Humidity 10% to 95%, non-condensing for monitor,
docking station, adapter cable, power supply
and removable battery
Storage and Transport
Temperature
-20 ºC to +60 ºC (-4 ºF to 140 ºF) for monitor,
docking station and adapter cable
-10 ºC to +60 ºC (14 ºF to 14 ºF) for BISx module
Storage and Transport
Pressure and Altitude
107kPa (430m [1410ft] below sea level) to 54kPa
(5000m [16400ft] above sea level) for monitor,
docking station and adapter cable
Storage and Transport
Humidity
10% to 95%, non-condensing, for monitor,
docking station, adapter cable, power supply
and removable battery
15% to 95% for BISx module
Start-up Time Up to 2 minutes
Ingress Protection marking Ingress Protection marking of the tablet, docking
station, adapter cable, and power supply is IPX2;
this is applicable only if the door on the right
edge of the tablet is closed.
IPX2 protection of the docking station is
applicable only if the tablet is connected to the
docking station.
For BISx and BISx4 modules, IP rating is IPX4