Administrator Mode
BIS™ Advance Monitor
7.3.2. Administrator Mode Actions
After the user enters a valid password and applies it, the software shall open the
Administrator Settings and Maintenance menu and enable the Administrator user to
perform the functions listed in this section.
Administrator Mode Settings
The user can modify defaults for the parameters listed in Table 19. Institutional Settings
on page 181 while in Administrator Mode (except for those items listed in that table
as not available in Administrator Mode).
To adjust Institutional settings, enter Administrator mode, and then click Main
Menu>Settings and Maintenance>Settings. Select the desired setting and
make changes as required.
The same settings can be adjusted by the user in standard mode, but in that case the
settings will revert to defaults when a new case is started or when the monitor is
turned off and on again. If changes are made and applied while in Administrator
Mode, the changes will remain as active defaults on the monitor (that is, the
institutional defaults) unless Factory Defaults are restored or changes are made in
Administrator mode, even if the monitor is turned off and on again.
In addition, some settings can be adjusted in Administrator Mode only. The settings
and actions that are available only in Administrator Mode are listed below. For other
settings, reference 4.6 Institutional Settings on page 180.
Table 26. Administrator Mode Settings
Feature Options Factory Default Option
Settings
EEG Testing (enable
display of EEG test
layout for EEG
testing) (when ON,
EEG Testing will
appear in Display
Settings list, until
the current case
ends or the
currently used
Enabled, disabled Disabled