Operator's Manual

Table Of Contents
Safety Information
BIS™ Advance Monitor
equipment may interfere with the proper function of the BIS™ Advance
monitor. Check for compatibility of equipment during patient setup.
Caution: The BISAdvance monitor complies with the electromagnetic
compatibility requirements of IEC 60601-1-2. Operation of this device may
affect or be affected by other equipment in the vicinity due to electromagnetic
interference (EMI). If this occurs:
• Increase separation between devices.
Re-orient device cabling.
• Plug devices into separate outlet circuit branches.
• Reference 7.7 Electromagnetic Compatibility Specifications on page 237.
Caution: When connecting or disconnecting the BISx module, take care not to
touch the exposed contacts of either connector. Damage due to electrostatic
discharge may result.
1.2.9. Notes
Note: Important: The BIS™ Advance monitor complies with the MDR- EU
Medical Device Regulation 2017/745 and applicable regulatory requirements
of the country distributed to and carry the CE Marking. Declarations of
Conformity provided upon request where appropriate.
Note: Read this entire manual carefully before using the monitor in a clinical
setting. The monitor shall be used only according to the instructions that
appear in this manual.
Note: Use of the host system with sensor along with a neurostimulator may
cause neurostimulator malfunction or diminished neurostimulator
effectiveness.
Note: The system provides protection against defibrillation, without loss of any
operator settings or stored data, and shall continue to perform its intended
functions within 30 seconds after exposure to defibrillation voltage, as
specified in IEC 80601-2-26 and IEC 60601-2-26.
Note: The battery pack supplied with this monitor should not be used with
other devices.
Note: Since the case storage capacity of the monitor is limited, it is
recommended to download cases when the case ends, to avoid accidental
erasure of case data.