Operator's Manual

Table Of Contents
Cleaning the BIS™ Advance Monitor
BIS™ Advance Monitor
Clean any spillage of blood or solutions on the surfaces of the monitor, docking
station, adapter cable or BISx module as soon as possible. Dried blood is very difficult
to remove. Use lint-free absorbent towels for spill cleanups. Dampen the towel with
detergent and lukewarm water to aid in cleaning. After cleaning, wipe the PIC
connector ends with alcohol and allow to dry completely. Residual moisture inside
the connector may affect BISx module performance.
The surfaces of the tablet, docking station, adapter cable, and power supply may also
be cleaned using the following materials:
Lint-free cloths for use with cleaning agent
Appropriate cleaning agent, as follows:
o Ethanol 70%
o Isopropanol 70%
o Hydrogen Peroxide solution 0.5-1% concentration
o Sodium Hypochlorite solution 6500 ppm (household bleach
diluted approximately 1:10)
6.4.2. Disinfecting the BIS™
Advance System
Use lint-free absorbent towels dampened with a 10% bleach solution, or a
commercial disinfectant (e.g., Lysol™* Professional Disinfectant Foam Cleaner Spray
or PDI Germicidal Disposable Wipes) to clean the surfaces of the monitor, docking
station, adapter cable and BISx module. After cleaning, dry all areas except the
monitor display screen (see 6.4.3 Cleaning the Monitor Display on page 210) with a
lint-free absorbent paper towel. Wipe the BISx module and PIC connector ends with
alcohol and allow to dry completely.
WARNING: Whenever an event such as spillage of blood or solutions occurs, re-
test leakage current before further use.
Caution: Do not autoclave the BISx module or monitor. Autoclaving will
seriously damage both components.
Caution: Avoid liquid ingress to the Patient Interface Cable (PIC), which is an
integral part of the BISx module. Contact of fluids with the PIC sensor
connectors can interfere with PIC performance.