BIS™ Advance Monitor
5. Data Storage, Transfer,
and Export
5.1. Cybersecurity and Data
Integrity
Caution: Do not connect any device to RS-232 (serial) port on the docking
station (reference 2.2.3 Docking Station on page 51) other than an approved
RS-232 cable. RS-232 cable is supplied by the customer.
Caution: Do not connect any USB cables to the USB port on the monitor
(reference 2.2.1 BIS™ Monitor on page 43) or to the USB port on the docking
station (reference 2.2.3 Docking Station on page 51). The USB ports are
intended only for the connection of a USB flash drive. USB flash drives that can
be used with the monitor must have a capacity of at least 0.5 GB.
Caution: The BIS™ Advance monitor has been validated for use only with the
BIS™ sensor. Electrodes or sensors that are not validated for use with the
monitor have not undergone a data integrity analysis and may potentially
impose a data security risk. For a list of sensors, reference Table 33. BIS™
Advance Monitor Components and Sensors on page 252.
Caution: For software updates, only use Medtronic's authorized software.
5.2. Monitor Data Memory
The following events are continuously logged in the monitor data memory with a
unique ID for easy identification. This data is used to generate the various logs
available for export, as described in 5.9 Logs Export on page 199.