Alarms and Messages
BIS™ Advance Monitor
1. Main Menu>Alarm Settings
2. Left Menu bar>Alarms button
3. Alarm Settings button that appears when a BISx module but not a
sensor is connected to the monitor, on the Connect sensor to start
monitoring window
4. The BIS target range can also be set using Home Screen>BIS Graph, Set
Range>BIS Range
In the first three cases, the same Alarm Settings window will open. The BIS range
setting choice will open a window that shows BIS range choices only.
Table 16. Alarm Limits Options
Alarm Range
Factory Default Alarm
Level
BIS range None
Low: 5 -95
High: 10-100
None; suggested level is
40-60
SR None, 1-100 None
SEF None
Low: 1-29
High: 2-30
None
Follow the steps listed below to change alarm settings:
1. By factory default, set when the monitor leaves the factory, all alarms are
set to OFF. If the alarms have been changed by the administrator in your
institution, some or all of the alarms may already be set to ON. You can still
change the alarm settings for your current monitoring session, but the
monitor will revert to its institutional default settings (or factory default
settings if institutional defaults have not been set) when a new case is
started or when the monitor is turned off and turned on again.
2. Attach the BISx module to the monitor.